- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE SUPPLY REGULATOR WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED FILAMENT REGULATOR FAILURE ERRORS DURING A PATIENT PROCEDURE. THIS EVENT MAY HAVE RESULTED IN AN ARO (ACCIDENTAL RADIATION OCCURRENCE). NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311515 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |