FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3915274 · Received May 27, 2014

Report

Report Number
1720753-2014-04497
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 28, 2014
Report Date
May 27, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE SUPPLY REGULATOR WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED FILAMENT REGULATOR FAILURE ERRORS DURING A PATIENT PROCEDURE. THIS EVENT MAY HAVE RESULTED IN AN ARO (ACCIDENTAL RADIATION OCCURRENCE). NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311515 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1