FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 3915228 · Received May 16, 2014

Report

Report Number
9615050-2014-03460
Event Type
Malfunction
Date Received
May 16, 2014
Date of Event
April 21, 2014
Report Date
April 21, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
2-1159-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND AT POWER UP THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THIS WAS DUE TO A DEPLETED LITHIUM BATTERY. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PROGRAMMING IN THE ANESTHESIA DEPARTMENT, THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM THE ANESTHESIA DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED "BROKEN". THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE ALARMED WITH AN 11/004 SERVICE ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
293572 GEMSTAR PAIN MGE SNG 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA