LOANER CORE IMPACTION DRILL

MDR report key: 3915166 · Received July 7, 2014

Report

Report Number
0001811755-2014-02406
Event Type
Malfunction
Date Received
July 7, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FAILURE FOR HEAT WAS CONFIRMED THROUGH FUNCTIONAL EVALUATION, DISASSEMBLY, AND VISUAL INSPECTION. THROUGH VISUAL INSPECTION THE TECHNICIAN CONFIRMED THE COUPLER, SOCKETS, AND BEARINGS WERE WORN AND CORRODED. THE DEVICE WAS REPAIRED AND PLACED BACK INTO STRYKER STOCK.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE LOANER CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.