FDA Adverse Event Malfunction Summary report: N

NEOPUFF INFANT RESUSCITATOR

MDR report key: 3915030 · Received July 6, 2014

Report

Report Number
9611451-2014-00602
Event Type
Malfunction
Date Received
July 6, 2014
Report Date
June 18, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTL
PMA / PMN Number
K892885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT NEOPUFF WAS RECEIVED AT OUR REGIONAL OFFICE IN (B)(6), WHERE IT WAS INSPECTED AND REPAIRED BY A TRAINED FISHER & PAYKEL HEALTHCARE SERVICE ENGINEER. OUR INVESTIGATION IS BASED ON PHOTOGRAPHS AND A SERVICE REPORT PROVIDED BY OUR (B)(6) OFFICE. RESULTS: VISUAL INSPECTION OF THE SUBJECT NEOPUFF REVEALED PHYSICAL DAMAGE TO THE GAS OUTLET PORT THAT APPEARED TO BE THE RESULT OF AN IMPACT TO THE DEVICE. THE UPPER END CAP WAS ALSO FOUND TO BE BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR A BROKEN OUTLET PORT FOR LOT 080129. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. IT MUST BE NOTED THAT THE SUBJECT NEOPUFF WAS SIX YEARS OLD. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT". THE NEOPUFF WAS REPAIRED AND WAS RETURNED TO THE CUSTOMER AFTER IT PASSED ALL PERFORMANCE AND SAFETY TESTS AS PER THE NEOPUFF TECHNICAL MANUAL.

Description of Event or Problem · 1

A DISTRIBUTOR IN (B)(6) REPORTED THAT A NEOPUFF INFANT RESUSCITATOR HAD A BROKEN PATIENT CONNECTOR AND REQUESTED A REPAIR.NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
393041 NEOPUFF INFANT RESUSCITATOR BTL BTL FISHER & PAYKEL HEALTHCARE LTD RD900 080129

Patients

Seq Age Sex Outcome Treatment
1