NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2014-00602
- Event Type
- Malfunction
- Date Received
- July 6, 2014
- Report Date
- June 18, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT NEOPUFF WAS RECEIVED AT OUR REGIONAL OFFICE IN (B)(6), WHERE IT WAS INSPECTED AND REPAIRED BY A TRAINED FISHER & PAYKEL HEALTHCARE SERVICE ENGINEER. OUR INVESTIGATION IS BASED ON PHOTOGRAPHS AND A SERVICE REPORT PROVIDED BY OUR (B)(6) OFFICE. RESULTS: VISUAL INSPECTION OF THE SUBJECT NEOPUFF REVEALED PHYSICAL DAMAGE TO THE GAS OUTLET PORT THAT APPEARED TO BE THE RESULT OF AN IMPACT TO THE DEVICE. THE UPPER END CAP WAS ALSO FOUND TO BE BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR A BROKEN OUTLET PORT FOR LOT 080129. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. IT MUST BE NOTED THAT THE SUBJECT NEOPUFF WAS SIX YEARS OLD. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT". THE NEOPUFF WAS REPAIRED AND WAS RETURNED TO THE CUSTOMER AFTER IT PASSED ALL PERFORMANCE AND SAFETY TESTS AS PER THE NEOPUFF TECHNICAL MANUAL.
A DISTRIBUTOR IN (B)(6) REPORTED THAT A NEOPUFF INFANT RESUSCITATOR HAD A BROKEN PATIENT CONNECTOR AND REQUESTED A REPAIR.NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393041 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 080129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |