FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3914867 · Received July 4, 2014

Report

Report Number
2955842-2014-04107
Event Type
Malfunction
Date Received
July 4, 2014
Date of Event
June 5, 2014
Report Date
June 13, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. NO OTHER DAMAGE OR WEAR MARKS WERE OBSERVED ON THE CLEVIS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT TIP WAS BROKEN - IT WOULD NOT OPEN. THE INSTRUMENT WAS NOT USED ON A PATIENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT, AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392817 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-07 N10140424 921