PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04105
- Event Type
- Malfunction
- Date Received
- July 4, 2014
- Date of Event
- June 6, 2014
- Report Date
- June 10, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. NO OTHER DAMAGE OR WEAR MARKS WERE OBSERVED ON THE CLEVIS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT AFTER A DA VINCI CHOLECYSTECTOMY PROCEDURE, FRAYED WIRES WERE IDENTIFIED AT THE TIP OF THE PK DISSECTING FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT, AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392831 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10130806 299 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |