MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2014-04098
- Event Type
- Malfunction
- Date Received
- July 4, 2014
- Date of Event
- May 4, 2014
- Report Date
- June 9, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENT'S WRIST. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT ANY DAMAGE. THE OTHER CABLES AT THE INSTRUMENT'S WRIST WERE INTACT AND UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE, IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, THE CABLE BROKE ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT AND THERE WERE NO LOOSE PIECES. THE INSTRUMENT HAD BEEN IN USE FOR APPROXIMATELY 15 MINUTES BEFORE THE CABLE BROKE AND THE INSTRUMENT HAD BEEN INSPECTED PRIOR TO USE. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392793 | MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-03 | M10140122 268 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |