VENTED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2014-00597
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 13, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4 FOR EVALUATION AND WAS VISUALLY INSPECTED. RESULTS: THE VISUAL INSPECTION REVEALED A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 131017. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT CAUSED THE REPORTED DAMAGE. HOWEVER THE DAMAGE APPEARS TO BE THE RESULT OF THE TUBE BEING PULLED AWAY FROM THE DOME, POSSIBLY DUE TO BEING CAUGHT OR PUT UNDER TENSION. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY PRODUCT FAILS THIS TEST, THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: SET APPROPRIATE VENTILATOR ALARMS. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT.
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THEY FOUND THE FEEDSET OF A MR290 HUMIDIFICATION CHAMBER WAS SNAPPED DURING SET-UP.THIS WAS DISCOVERED BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391807 | VENTED HUMIDIFICATION CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 1310170105 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |