FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 3914675 · Received July 3, 2014

Report

Report Number
9611451-2014-00597
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 11, 2014
Report Date
June 13, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4 FOR EVALUATION AND WAS VISUALLY INSPECTED. RESULTS: THE VISUAL INSPECTION REVEALED A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 131017. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT CAUSED THE REPORTED DAMAGE. HOWEVER THE DAMAGE APPEARS TO BE THE RESULT OF THE TUBE BEING PULLED AWAY FROM THE DOME, POSSIBLY DUE TO BEING CAUGHT OR PUT UNDER TENSION. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY PRODUCT FAILS THIS TEST, THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: SET APPROPRIATE VENTILATOR ALARMS. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THEY FOUND THE FEEDSET OF A MR290 HUMIDIFICATION CHAMBER WAS SNAPPED DURING SET-UP.THIS WAS DISCOVERED BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391807 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1310170105

Patients

Seq Age Sex Outcome Treatment
1