FDA Adverse Event Malfunction Summary report: N

STRATA 2 VALVE, SMALL

MDR report key: 3914511 · Received July 3, 2014

Report

Report Number
2021898-2014-00227
Event Type
Malfunction
Date Received
July 3, 2014
Report Date
June 6, 2014
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K012052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS UNAVAILABLE FOR RETURN. THEREFORE AN EVALUATION OF DEVICE PERFORMANCE WAS NOT POSSIBLE. IT IS UNKNOWN IF THE PATIENT¿S VALVE SITE HAD BEEN EXPOSED TO STRONG MAGNETS. STRATA PRODUCTS ARE DESIGNED TO BE ADJUSTED BY A STRONG MAGNET. INADVERTENT ADJUSTMENTS BY EXTERNAL MAGNETS ARE A KNOWN COMPLICATION WITH ADJUSTABLE VALVES. A REVIEW OF THE MANUFACTURING RECORDS WAS NOT POSSIBLE AS NO LOT NUMBER WAS PROVIDED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT THE PHYSICIAN WAS EXPERIENCING SPONTANEOUS ADJUSTMENTS WITH THE STRATA VALVES. ACCORDING TO THE REPORT, HE HAD SEEN BOTH INCREASE AND DECREASE ADJUSTMENTS IN THE VALVES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389796 STRATA 2 VALVE, SMALL SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS JXG MEDTRONIC NEUROSURGERY UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1