COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2014-00590
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Report Date
- June 5, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT IW933 COSYCOT INFANT WARMER WAS INSPECTED BY A TRAINED FPH SERVICE TECHNICIAN AT OUR SERVICE CENTRE IN (B)(4). RESULTS: DURING TESTING IT WAS NOTED THAT THE UNIT'S POWER FAIL ALARM DID NOT FUNCTION WHEN THE UNIT WAS DISCONNECTED FROM POWER. THE ROOT CAUSE WAS IDENTIFIED TO BE THE FAILURE OF A CAPACITOR ON THE PCB BOARD. THE UPPER HEAD CASE OF THE INFANT WARMER WAS ALSO OBSERVED AS BEING CRACKED. CONCLUSION: THE SUBJECT IW933 COSYCOT INFANT WARMER WAS RELEASED FOR DISTRIBUTION IN 2004. IT IS LIKELY THAT THE REPLACEABLE SUPER CAPACITOR ON THE PCB BOARD WORE OUT OVER TIME. THE SUPER CAPACITOR'S FUNCTION IS TO STORE SUFFICIENT ELECTRICAL CHARGE TO POWER THE WARMER'S VISUAL AND AUDIBLE ALARMS IN THE EVENT OF A FAILURE OF MAINS POWER. PART OF FISHER & PAYKEL HEALTHCARE'S QUALITY CONTROL PROCESS INVOLVES TESTING THE POWER FAILURE ALARM OF EVERY WARMER ON THE PRODUCTION LINE FOR FUNCTIONALITY PRIOR TO DISTRIBUTION. THE DEVICE TECHNICAL/SERVICE MANUAL CONTAINS A CHECKLIST WHICH SPECIFIES THAT USERS PERFORM SAFETY, PERFORMANCE AND FUNCTIONAL CHECKS INCLUDING THE POWER FAIL ALARM AT LEAST ONCE A YEAR. INCLUDED IN THE WARMER'S MAINTENANCE CHECKLIST IS A TEST OF THE SUPER CAPACITOR TO ENSURE IT IS OPERATING WITHIN SPECIFICATIONS. THE REPORTED MALFUNCTION WAS DISCOVERED DURING A MAINTENANCE CHECK WITH NO PATIENT INVOLVEMENT. WITH REGARDS TO THE CRACKED HEAD, IT IS LIKELY THAT THE CRACKING AND CRAZING OF THE COMPLAINT WARMER HEAD IS RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE HEATER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE HEATER HEAD. THE CRACKING COULD ALSO BE THE RESULT OF IMPACT DAMAGE AND WEAR AND TEAR WHILE MOVING THE WARMER BETWEEN WARDS. THE INFANT WARMER SERVICE MANUAL FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE; CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES. THE INFANT WARMER WAS REPAIRED AND A NEW CAPACITOR WAS FITTED TO THE PCB. THE SUBJECT INFANT WARMER WAS RETURNED TO THE CUSTOMER AFTER PASSING ALL THE NECESSARY SAFETY AND PERFORMANCE TESTS.
(B)(4). THE SERVICE REPORT RELATED TO THE COMPLAINT DEVICE HAS JUST BEEN RECEIVED AT FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A SERVICE CONTRACTOR REPORTED ON BEHALF OF A HOSPITAL IN (B)(6) THAT AN IW933 COSYCOT INFANT WARMER'S POWER FAIL ALARM DOES NOT WORK. THE UPPER CASE WAS ALSO OBSERVED AS BEING CRACKED.THIS WAS FOUND DURING SERVICING.
A SERVICE CONTRACTOR REPORTED ON BEHALF OF A HOSPITAL IN (B)(6) THAT AN IW933 COSYCOT INFANT WARMER'S POWER FAIL ALARM DOES NOT WORK. THE UPPER CASE WAS ALSO OBSERVED AS BEING CRACKED. THIS WAS FOUND DURING SERVICING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390275 | COSYCOT INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW933 | 040622 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |