SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Report
- Report Number
- 1416980-2014-21609
- Event Type
- Injury
- Date Received
- July 3, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 11, 2014
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE CAUSE OF THIS PERITONITIS WAS USE ERROR REPORTED TO BE DUE TO A BREAK IN ASEPTIC TECHNIQUE. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). ON AN UNREPORTED DATE, THE PATIENT WAS RECOVERED FROM THE PERITONITIS EVENT (PREVIOUSLY REPORTED AS RECOVERING). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THIS IS A REPORT OF A PATIENT WHO EXPERIENCED A BREACH IN ASEPTIC TECHNIQUE RESULTING IN PERITONITIS COINCIDENT WITH PERITONEAL DIALYSIS (PD) THERAPY. THE BREACH IN ASEPTIC TECHNIQUE WAS FURTHER DESCRIBED AS THE PATIENT DID NOT WEAR A MASK AND DID NOT CLEAN THE AREA BEFORE STARTING PD THERAPY. THE PATIENT EXPERIENCED PERITONITIS AND WAS HOSPITALIZED ON THE SAME DAY FOR THE EVENT. THE PATIENT WAS TREATED WITH INTRAPERITONEAL (IP) FORTUM (1G, FREQUENCY NOT REPORTED) AND IP VANCOMYCIN (1G, FREQUENCY NOT REPORTED) FOR PERITONITIS. DIANEAL THERAPIES WERE ONGOING. AT THE TIME OF THIS REPORT, THE PATIENT WAS RECOVERING. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390274 | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 5 YR | Hospitalization| R | DIANEAL 2.5% ULTRABAG THERAPY| DIANEAL 1.5% ULTRABAG THERAPY |