FDA Adverse Event Malfunction Summary report: N

7.5CM SHORT ANGLE ATTACHMENT

MDR report key: 3914456 · Received July 3, 2014

Report

Report Number
1045834-2014-12376
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
April 1, 2014
Report Date
April 30, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. VISUAL ASSESSMENT WAS PERFORMED AND FOUND A BUSHING FALLING WHICH WHAT WAS MISTAKENLY IDENTIFIED AS A BEARING FALLING OUT BY THE CUSTOMER. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DUE TO NORMAL WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE BEARINGS ON THE ATTACHMENT DEVICE CAME OUT. THERE WERE NO DELAYS TO THE SCHEDULED SURGICAL PROCEDURE AS A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE REPORTED EVENT WAS DISCOVERED ON (B)(6) 2014; HOWEVER, THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390223 7.5CM SHORT ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1