HOMECHOICE PRO

MDR report key: 3914394 · Received July 3, 2014

Report

Report Number
1416980-2014-21601
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 1, 2014
Report Date
June 11, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES VISUAL INSPECTION, AND FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 20:52:36. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1820ML, INDICATING THE HOME PATIENT (HP) DRAINED 1820ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2600ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389994 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 61 YR