FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3914327 · Received July 3, 2014

Report

Report Number
1416980-2014-21573
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 23, 2013
Report Date
June 11, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. THE CAUSE WAS DETERMINED TO BE UNEXPECTED HIGH UF FOR THERAPY COMPARED TO OTHER THERAPIES. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT(S) WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 23:56:42. DURING NIGHT DRAIN CYCLE FIVE, THE HOME PATIENT (HP) DRAINED MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME. NO ADDITIONAL INFORMATION IS AVAILABLE.