FDA Adverse Event Malfunction Summary report: N

AU2700 CHEMISTRY ANALYZER

MDR report key: 3914277 · Received July 3, 2014

Report

Report Number
9612296-2014-00106
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 5, 2014
Report Date
June 6, 2014
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K003721
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE REPLACED THE FIBER OPTIC LAMP BUNDLE, WHICH RESTORED THE INSTRUMENT TO FUNCTIONALITY. THE FSE PERFORMED AN INSTRUMENT PHOTOCAL AND THERE WERE NO FURTHER ISSUES REPORTED. ASSOCIATED MDR: 9612296-2014-00103. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN AU2700 CHEMISTRY ANALYZER GENERATED ELEVATED LIPASE, CREATINE KINASE (CK), AND AMYLASE RESULTS. IN ADDITION, SEVERAL INNER CUVETTES FAILED THE MEAN CHECK DURING INSTRUMENT PHOTOCAL. THE CUSTOMER PROVIDED DATA FOR THREE PATIENT SAMPLES, WHICH SHOWED ELEVATED LIPASE OR CK RESULTS; THIS MDR REPORTS THE PATIENT RESULTS GENERATED ON (B)(6) 2014. ON (B)(6) 2014, THE INSTRUMENT GENERATED AN ELEVATED LIPASE RESULT FOR ONE PATIENT, WHICH WAS FLAGGED TO INDICATE THE RESULT WAS HIGHER THAN DYNAMIC RANGE AND REPEAT RANGE. REPEAT ANALYSIS ON THE SAME INSTRUMENT PRODUCED A NEGATIVE LIPASE RESULT, WHICH WAS FLAGGED TO INDICATE THE RESULT WAS LOWER THAN DYNAMIC RANGE AND OUTSIDE THE PANIC VALUE RANGE. THERE WERE NO ERRONEOUS PATIENT RESULTS RELEASED FROM THE LABORATORY AND THERE WERE NO REPORTS OF CHANGE TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391428 AU2700 CHEMISTRY ANALYZER ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER AU2700 NA

Patients

Seq Age Sex Outcome Treatment
1