FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3914158 · Received July 3, 2014

Report

Report Number
1416980-2014-21545
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
March 1, 2014
Report Date
June 11, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. A REVIEW OF THE EVENT HISTORY LOG DID NOTE IDENTIFICATION OF THE REPORTED ISSUE, BUT NO NONCONFORMING PRODUCT WAS IDENTIFIED RELATED TO THE REPORTED ISSUE. THE DIRECT CAUSE WAS DETERMINED TO BE USE ERROR AND TIDAL TOTAL UF REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY ADDITIONAL RELEVANT INFORMATION IS OBTAINED OR ANY OTHER ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT(S) WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 00:33:50. DURING NIGHT DRAIN CYCLE EIGHT, THE PATIENT'S ULTRAFILTRATION READING WAS 1924ML, INDICATING THE HOME PATIENT (HP) DRAINED 1484ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391514 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1