FDA Adverse Event Malfunction Summary report: N

TENACULUM FORCEPS INSTRUMENT

MDR report key: 3914112 · Received July 3, 2014

Report

Report Number
2955842-2014-04088
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
May 29, 2014
Report Date
June 9, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER REPORTED COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE INSTRUMENT PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS MISSING FROM THE CLEVIS. CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER IF THE BROKEN CABLE WERE TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED DURING DA VINCI HYSTERECTOMY PROCEDURE THAT THE TENACULUM FORCEPS INSTRUMENT'S CABLE BROKE/SNAPPED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391845 TENACULUM FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420207-07 M10140124 647