FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3914051 · Received July 3, 2014

Report

Report Number
2531779-2014-18997
Event Type
Injury
Date Received
July 3, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 08/06/2014-PRODUCT ANALYSIS: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP¿S BLACK BOX AND HISTORY WERE REVIEWED. THE TOTAL DAILY DOSE HISTORY CORRECTLY REFLECTED THE PROGRAMMED BASAL RATES. THE PRIME HISTORY REVEALED SMALL PRIME VOLUMES DURING CARTRIDGE CHANGES. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. A DELIVERY ACCURACY TEST SHOWED THE PUMP¿S DELIVERY ACCURACY WAS WITHIN SPECIFICATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, IT WAS REPORTED THE PATIENT¿S BLOOD GLUCOSE WAS ABOVE 600 MG/DL FROM (B)(6) 2014 TO (B)(6) 2014 WITH LEG MUSCLE WEAKNESS. IT WAS REPORTED THE PUMP SETTINGS HAD NOT BEEN CHANGED RECENTLY AND THE PATIENT SELF-TREATED WITH INSULIN DELIVERY THROUGH ANOTHER INSULIN PUMP. TROUBLESHOOTING COULD NOT BE COMPLETED. THIS COMPLAINT IS BEING REPORTED SINCE THE ALLEGED ADVERSE EVENT WAS ASSOCIATED WITH AN UNKNOWN ALLEGED INACCURATE DELIVERY ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391768 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 64 YR Life Threatening