MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2014-04085
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 30, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO REPLICATE THE CUSTOMER REPORTED COMPLAINT. THE INSTRUMENT WAS PLACED ON AN IN HOUSE SYSTEM AND DRIVEN AND PASSED RECOGNITION AND ENGAGEMENT TESTING. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE SCISSORS OPENED AND CLOSED PROPERLY. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE INSTRUMENT'S TUBE EXTENSION HAD PAD PRINTING REMOVED. THE PAD PRINTING HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL MISSING. THE SCRATCHED AREA MEASURED APPROXIMATELY .143 X .362. FAILURE ANALYSIS CONCLUDED THAT THE PAD PRINTING REMOVAL WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE PAD PRINTING REMOVAL FOUND DURING FAILURE ANALYSIS IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY SURGICAL PROCEDURE, IT WAS NOTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT'S JAW WAS NOT ROTATING WELL. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 389757 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-12 | M10131212 761 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |