PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04087
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- May 28, 2014
- Report Date
- June 11, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE PITCH CABLE WAS BROKEN AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY WEAR. THE BROKEN STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE FOLLOWING ADDITIONAL DAMAGES TO THE INSTRUMENT: ONE GRIP WAS BENT, CAUSING SIDE-TO-SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .007 OFFSET AT THE TIPS. THE BENT GRIP DID NOT EXHIBIT SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT, LIKELY CAUSE OF BENDING REMAINS OVERLOADING AT THE TIP. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE AND/OR THE TUBE ABRASIONS, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE PK DISSECTING FORCEPS INSTRUMENT WIRE WAS NOTED TO BE FRAYED DURING A DA VINCI SURGICAL PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 389639 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | N10140227 432 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |