ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2014-19115
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Report Date
- June 12, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).
(B)(4). DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY WAS FOUND TO BE BLANK WHEN THE PUMP WAS BOOTED. A SINGLE COMPONENT FAILURE WAS FOUND. THE DISPLAY WAS REPLACED AND THE PUMP WAS REBOOTED AND DISPLAYED THE VERIFY SCREEN. UNRELATED TO THE COMPLAINT, THE AUDIO FEATURE WAS FOUND TO BE FUNCTIONING INTERMITTENTLY DUE TO A LOOSE CONNECTION AT THE PIEZO CONTACTS.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.