ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-18924
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Report Date
- June 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT ONE REPLACE BATTERY ALARM AND ONE CALL SERVICE ALARM WAS RECORDED ON (B)(6) 2014. DURING TESTING, THE PUMP POWERED ON WITH AUDIO TONES AND VIBRATIONS FUNCTIONAL. THE DISPLAY WAS FOUND TO BE BLANK. THE PUMP CASE WAS REMOVED AND EVIDENCE OF MOISTURE CONTAMINATION WAS FOUND INSIDE THE PUMP. THE CALL SERVICE ALARM ISSUE WAS NOT ABLE TO BE ADEQUATELY INVESTIGATED DUE TO THE BLANK DISPLAY.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THERE WAS NO FURTHER INFORMATION AVAILABLE REGARDING THE COMPLAINT AVAILABLE AT THIS TIME; IF FURTHER INFORMATION IS PROVIDED A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391792 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |