FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3913399 · Received July 3, 2014

Report

Report Number
2531779-2014-18932
Event Type
Malfunction
Date Received
July 3, 2014
Report Date
June 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: BLACK BOX AND HISTORY RECORDED 064-0001, 070-0001, AND 078-0004 ALARMS. DURING INVESTIGATION RECEIVED A CS 076 ALARM (REWIND TOLERANCE ERROR). OPENED PUMP AND REMOVED FROM CASE. DISASSEMBLED MOTOR DRIVE ASSEMBLY. TESTED MOTOR ON A 5 VOLT POWER SUPPLY. ABLE TO EASILY STALL MOTOR. UNABLE TO COMPLETE FAILED STEPS DUE TO MOTOR FAILURE. UNRELATED TO THE COMPLAINT, THE DISPLAY WAS FADED AND PINK.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A MOTOR ISSUE. THE PISTON STOPPED MOVING COMPLETELY. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391652 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1