FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3913351 · Received May 14, 2014

Report

Report Number
2518422-2014-00809
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 17, 2014
Report Date
April 17, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SENSOR BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. OTHER - DEVICE HAS BEEN EVALUATED BUT NOT REPAIRED, PENDING CUSTOMER APPROVAL OF THE ESTIMATE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A VENTILATOR FAILED A STEP DURING TESTING. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288061 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054096

Patients

Seq Age Sex Outcome Treatment
1