NC TREK CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2014-04352
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- April 29, 2014
- Report Date
- April 29, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- K110134
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED FAULTY/BROKEN DEVICE WAS CONFIRMED AS THERE WAS A TEAR IN THE DISTAL SHAFT. BASED ON VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR TORN SHAFT REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT THE NC TREK BALLOON WAS NOTED TO BE FAULTY/BROKEN PRIOR TO USE. THE DEVICE WAS NOT USED ON THE PATIENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY OF PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. RETURNED DEVICE ANALYSIS REVEALED A TEAR IN THE DISTAL SHAFT, 2 CM PROXIMAL TO THE PROXIMAL SEAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390115 | NC TREK CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 40209G1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |