FDA Adverse Event Malfunction Summary report: N

NC TREK CORONARY DILATATION CATHETER

MDR report key: 3913267 · Received July 3, 2014

Report

Report Number
2024168-2014-04352
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
April 29, 2014
Report Date
April 29, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110134
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED FAULTY/BROKEN DEVICE WAS CONFIRMED AS THERE WAS A TEAR IN THE DISTAL SHAFT. BASED ON VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR TORN SHAFT REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NC TREK BALLOON WAS NOTED TO BE FAULTY/BROKEN PRIOR TO USE. THE DEVICE WAS NOT USED ON THE PATIENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY OF PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. RETURNED DEVICE ANALYSIS REVEALED A TEAR IN THE DISTAL SHAFT, 2 CM PROXIMAL TO THE PROXIMAL SEAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390115 NC TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 40209G1

Patients

Seq Age Sex Outcome Treatment
1