FDA Adverse Event Malfunction Summary report: N

SAGITTAL SAW ATTACHMENT

MDR report key: 3913257 · Received July 3, 2014

Report

Report Number
8030965-2014-10351
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 13, 2014
Report Date
June 19, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. A FUNCTIONAL ASSESSMENT WAS PERFORMED WHICH FOUND THAT THE BEARINGS AND SHAFTS WERE DAMAGED. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE MOST LIKELY DUE TO NORMAL USE OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FEMORAL HEAD OSTEOTOMY (FHO) SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE SAGITTAL SAW ATTACHMENT DEVICE DID NOT WORK. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389922 SAGITTAL SAW ATTACHMENT INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF 13673

Patients

Seq Age Sex Outcome Treatment
1