SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2014-03124
- Event Type
- Malfunction
- Date Received
- January 14, 2014
- Date of Event
- December 1, 2013
- Report Date
- December 18, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- 1314492-080712-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND TO BE OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM, WHICH WAS REPRODUCED. IMPROPER KEYPAD FUNCTION WAS CONFIRMED AND WAS DETERMINED TO BE CAUSED BY A FAILED KEYPAD. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #7 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #7 KEY IS PRESSED 17 WILL BE DISPLAYED"). WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AS WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH. THE FAILED KEYPAD WAS REPLACED. BAXTER HAS INITIATED A CAPA TO FURTHER THE REPORTED EVENT.
IT WAS REPORTED THAT A SPECTRUM PUMP'S KEYBOARD WAS NOT FUNCTIONING PROPERLY. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.