ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2014-04635
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 13, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULT SHOWED THAT THE ER320 INSTRUMENT WAS RECEIVED EMPTY AND WITH THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. IT IS KNOWN FROM THE HISTORY OF THE DEVICE THAT THE CONDITION OF THE JAWS MAY LEAD DROPPING/EJECTED CLIPS. DUE TO THE CONDITION OF THE RETURNED DEVICE IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WERE MALFORMED. UNKNOWN AT WHAT FIRING. THEN THE CLIPS STARTED FALLING OUT OF THE DEVICE. IT IS UNKNOWN IF THEY FELL IN THE PATIENT. THE SURGEON WAS TORQUING THE DEVICE DURING FIRING. A SECOND DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390886 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |