FDA Adverse Event Malfunction Summary report: N

INTERSTIM

MDR report key: 3912747 · Received July 3, 2014

Report

Report Number
9614453-2014-01528
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
May 1, 2014
Report Date
June 11, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
EZW
PMA / PMN Number
P970004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED A JOLT WHILE SITTING IN A HOTEL ROOM. THE PATIENT WAS UNSURE IF IT WAS HER SPINE OR THE IMPLANTABLE NEUROSTIMULATOR (INS). THE PATIENT TURNED THE INS OFF AND THE ISSUE WENT AWAY. THE INCIDENT HAPPENED TWICE WHILE IN THE HOTEL ROOM WITH NO OTHER EQUIPMENT AROUND. IT WAS ALSO SAID TO HAVE HAPPENED ONCE WHEN THE PATIENT WAS GETTING HER NAILS DONE. THE SHOCKING WAS SAID TO HAVE STARTED IN (B)(6) 2014. THE REPORTER STATED THAT MANUFACTURER REPRESENTATIVE DID NOT THINK THE ISSUE WAS POSITION RELATED, EVEN THOUGH THE SHOCKS OCCURRED WHILE SITTING. THE PATIENT ALSO MENTIONED BEING SORE AND FELT STIMULATION ANTERIOR TO THE ANUS; HOWEVER, THE JOLT WAS MORE TOWARD THE TAILBONE OR INS. THE PATIENT WAS CURRENTLY SET AT C+, 1- AT 3.0 VOLTS. THE PATIENT DENIED ANY FALLS OR TRAUMA ASSOCIATED WITH THE EVENT. IMPEDANCES WERE TESTED AND WERE UNREMARKABLE (SEE RESULTS BELOW). THE REPORTER WAS TO TRY BIPOLAR PROGRAMMING TO SEE IF PATIENT COULD USE A LOWER AMPLITUDE. IT WAS ALSO REPORTED THEN WHEN THE MANUFACTURER REPRESENTATIVE INTERROGATED THE INS TODAY WITH THE CLINICIAN PROGRAMMER, A MESSAGE WAS RECEIVED STATING THAT THE INS WAS PROGRAMMED BY AN INCOMPATIBLE PROGRAMMER. THE MANUFACTURER REPRESENTATIVE THEN WENT AHEAD AND PROGRAMMED THE INS. IT WAS STATED THAT THE INS HAD NOT BEEN INTERROGATED SINCE IMPLANT APPROXIMATELY 7 YEARS AGO. IT WAS DISCUSSED THAT THE ERROR MAY BE DUE TO THE USE OF A DIFFERENT DATA CARD SINCE THE LAST INTERROGATION. IT WAS ALSO STATED THAT THE PATIENT PROGRAMMER SHOWED A LOW BATTERY, BUT THE CLINICIAN PROGRAMMER SHOWED A LONGEVITY OF 102 MONTHS FROM THE DATE OF IMPLANT. AFTER THIS, THE PATIENT PROGRAMMER BATTERY WAS OK. IT WAS NOTED THAT THE PATIENT KEPT THE INS OFF FOR MOST OF THE DAY. THE PATIENT USED THERAPY FOR 30 MINUTES, 3 TIMES A DAY AND HAD DONE SO SINCE IMPLANT. [IMPEDANCE TESTING (ALL RESULTS ARE IN OHMS): C/0= 616, C/1=556, C/2=556, C/3=1111, 0/1=1111, 0/2=1111, 0/3=1111, 1/3-1111, 1/3=1111, 2/3=1111 AND 1/2=1077; THE IMPEDANCE WAS RETESTED AT 2.0 VOLTS AND PULSE WIDTH OF 330 MICROSECONDS: C/0= 724, C/1=691, C/2=691, C/3=872, 0/1=872, 0/2=1171, 0/3=1184, AND 1/2=865, 1/3=1171, AND 2/3=872] IT WAS REPORTED 6 DAYS LATER THAT THE CAUSE OF THE EVENT WAS NOT DETERMINED. IT WAS SUSPECTED THAT THE UNIPOLAR VECTOR MIGHT BE THE SOURCE (USING A UNIPOLAR ELECTRODE COMBINATION), BUT IT WOULD INVESTIGATED FURTHER AT THE NEXT VISIT. IT WAS STATED THAT THERAPY WAS EXPECTED TO BE OK AS A BIPOLAR VECTOR INDICATED APPROPRIATE STIMULATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389614 INTERSTIM STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE EZW IPG MFG SWITZERLAND 3023

Patients

Seq Age Sex Outcome Treatment
1