SYMBIQ DUAL CHANNEL

MDR report key: 3912713 · Received May 14, 2014

Report

Report Number
9615050-2014-03428
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
February 26, 2014
Report Date
April 16, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. FURTHER TESTING FOUND FLUID INGRESS ON THE BOTTOM OF THE TOUCHSCREEN WHICH WAS THE PROBABLE CAUSE OF THE NON RESPONSIVE TOUCHSCREEN. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. THERE WAS NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
287953 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA

Patients

Seq Age Sex Outcome Treatment
1 NA