FDA Adverse Event Malfunction Summary report: N

PLUM A + DRIVER NEW

MDR report key: 3912660 · Received May 14, 2014

Report

Report Number
9615050-2014-03409
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 17, 2014
Report Date
April 25, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042801
Removal / Correction Number
FA210-07
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE PASSED TESTING. AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE WAS NOT NOTED IN THE DEVICE HISTORY BUT WAS NOT DUPLICATED DURING TESTING. ALTHOUGH THE DEVICE PASSED TESTING, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE WAS NOTED. THIS ERROR OCCURS WHEN THE BATTERY CANNOT BE FULLY CHARGED WITHIN 8 HRS. THE DEVICE WAS RETURNED TO THE BIOMED DEPT FOR A REPORT OF MALFUNCTION. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288261 PLUM A + DRIVER NEW 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE, LIST #UNK, SN UNK