FDA Adverse Event Injury Summary report: N

ACRYSOF IQ TORIC

MDR report key: 3912647 · Received June 13, 2014

Report

Report Number
1119421-2014-00452
Event Type
Injury
Date Received
June 13, 2014
Date of Event
March 6, 2014
Report Date
May 19, 2014
Manufacturer
ALCON RESEARCH, LTD/ HUNTINGTON
Product Code
HQL
PMA / PMN Number
P930014
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASE CRITERIA. ADDITIONAL INFORMATION WAS PROVIDED, WHICH INDICATED AN APPROVED CARTRIDGE AS USED BUT AN UNAPPROVED HANDPIECE WAS USED WITH THE VISCOELASTIC. NOT ENOUGH INFORMATION WAS PROVIDED TO CONDUCT A REVIEW ON THE CARTRIDGE LOT. THE CARTRIDGE DESIGN WILL NOT WORK WITH THE HANDPIECE USED. THE PRODUCT INVESTIGATION COULD NOT IDENTIFY A ROOT CAUSE. ACCOUNT USED AN UNAPPROVED HANDPIECE. THE USE OF UNAPPROVED PRODUCTS MAY RESULT IN LENS DELIVERY DIFFICULTIES OR LENS DAMAGE. THERE HAVE BEEN NO OTHER COMPLAINTS REPORTED IN THE LOT NUMBER. ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION BY PHONE, FAX, AND MAIL. A COMPLETED QUESTIONNAIRE WAS RECEIVED ON 05/23/2014. (B)(4).

Description of Event or Problem · 1

A CLINICAL MANAGER REPORTED FOLLOWING INTRAOCULAR LENS (IOL) IMPLANT SURGERY, A PATIENT HAD AN UNEXPECTED REFRACTIVE ERROR. THE LENS WAS EXCHANGED. ADDITIONAL INFORMATION HAS BEEN REQUESTED. IN A FOLLOW UP, THE SURGEON REPORTED MULTIPLE EYE DROP MEDICATIONS WERE USED TO TREAT THE PATIENT DURING THE EVENT. THE PATIENT'S SYMPTOMS RESOLVED WITH THE LENS EXCHANGE. AN UNAPPROVED HANDPIECE WAS USED DURING THE INITIAL IMPLANT SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349124 ACRYSOF IQ TORIC INTRAOCULAR LENS HQL ALCON RESEARCH, LTD/ HUNTINGTON SN6AT8 12179409

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention SHUGARCAINW| DUOVISC| MONARCH II HANDPIECE| VIGAMOX| EPI 1:1000| MONARCH D CARTRIDGE| BSS