FINELINE II
Report
- Report Number
- 2124215-2014-12913
- Event Type
- Injury
- Date Received
- July 3, 2014
- Date of Event
- June 8, 2014
- Report Date
- August 26, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- DTB
- PMA / PMN Number
- P960004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE LEAD WAS SURGICALLY ABANDONED. AS THE LEAD WILL NOT BE RETURNED FOR ANALYSIS, THE CLINICAL OBSERVATIONS CANNOT BE CONFIRMED.
(B)(4)
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
ADDITIONAL INFORMATION RECEIVED THAT THIS RIGHT ATRIAL (RA) LEAD WAS ABANDONED SURGICALLY.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD WAS BANGING ON THE RIGHT VENTRICULAR (RV) LEAD. ADDITIONAL INFORMATION RECEIVED FROM THE FIELD REPRESENTATIVE INDICATED THAT THE RA LEAD HAD DISLODGED. FURTHERMORE, THE FIELD REPRESENTATIVE DID NOT KNOW IF THE POSITION WAS CONFIRMED THROUGH X-RAY. THE RA WAS REPOSITIONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD MOVED AND WAS IN CONTACT WITH THE RIGHT VENTRICULAR (RV) LEAD. ADDITIONAL INFORMATION RECEIVED FROM THE FIELD REPRESENTATIVE INDICATED THAT THE RA LEAD HAD DISLODGED. FURTHERMORE, THE FIELD REPRESENTATIVE DID NOT KNOW IF THE POSITION WAS CONFIRMED THROUGH X-RAY. THE RA WAS THEN REPOSITIONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 389629 | FINELINE II | IMPLANTABLE LEAD | DTB | CPI - DEL CARIBE | 4469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| L| R | 4469| N161| 4549| 0180 |