FDA Adverse Event Malfunction Summary report: N

ZIMMER SKIN GRAFT MESHER

MDR report key: 3912549 · Received May 14, 2014

Report

Report Number
1526350-2014-00356
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 1, 2014
Report Date
April 15, 2014
Manufacturer
ZIMMER SURGICAL
Product Code
FZW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR FOR REPAIR AND EVALUATION. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 9/8/2010 AND WAS LAST REPAIRED ON 1/7/2014 FOR A RELATED EVENT. EVALUATION OF THE DEVICE OBSERVED THAT THE COMB, ROLLER AND ROLLER GEAR WERE DAMAGED. IT WAS ALSO NOTED THAT BOTH CUTTER'S GEAR TEETH WERE DAMAGED. PRIOR TO REPAIR, BOTH THE TEST MESH AND THE CALIBRATION CHECK COULD NOT BE PERFORMED DUE TO THE DAMAGED COMB. BOTH CUTTERS DID NOT PASS CUTTER EVALUATION AND WERE RETURNED TO THE CUSTOMER AS NON-REPAIRABLE. REPAIR OF THE DEVICE INCLUDED REPLACEMENT OF THE COMB, ROLLER AND ROLLER GEAR, CUTTER GEARS, COLLARS AND BUSHINGS. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. THE CUTTERS WERE RETURNED TO THE CUSTOMER AS NON-REPAIRABLE. THE CAUSE OF THE REPORTED ISSUE REVEALED THAT IMPROPER HANDLING BY THE USER MOST LIKELY CAUSED THE DAMAGE TO THE COMB, CUTTER AND ROLLER GEARS AND CUTTERS, WHICH MOST LIKELY CONTRIBUTED TO THE CUSTOMER'S REPORTED EVENT REGARDING THE DAMAGED GEARS AND THE DEVICE PRODUCING AN INCOMPLETE MESH. IT SHOULD BE NOTED THAT BOTH CUTTERS WERE PREVIOUSLY RETURNED TO THE CUSTOMER AS NON-REPAIRABLE. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT ALL GEARS OF THE ZIMMER SKIN GRAFT MESHER WERE DAMAGED PRODUCING AN INCOMPLETE MESH. THE FACILITY REPORTING THE EVENT IS A CADAVER TISSUE BANK. AS SUCH, THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288206 ZIMMER SKIN GRAFT MESHER ZIMMER SKIN GRAFT MESHER FZW ZIMMER SURGICAL NA NA

Patients

Seq Age Sex Outcome Treatment
1 ZIMMER SKIN GRAFT MESHER CUTTER 2:1 RATIO 911046| ZIMMER SKIN GRAFT MESHER CUTTER 1.5:1,| RATIO 1203008