FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 3912457 · Received May 14, 2014

Report

Report Number
2028159-2014-00824
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 16, 2014
Report Date
April 16, 2014
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND WAS NOT ABLE TO REPLICATE THE REPORTED EVENT. THE COMPANY REPRESENTATIVE REPLACED THE POWER SUPPLY AND POWER DISTRIBUTION PRINTED CIRCUIT BOARD (PCB) FOR DIAGNOSTIC PURPOSES. THE COMPANY REPRESENTATIVE THEN TESTED THE SYSTEM. THE SYSTEM MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVAL AND NO ADDITIONAL INFO PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS INDICATED 1 ADDITIONAL SIMILAR REPORT FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE SYSTEM SHUT DOWN ON ITS OWN BETWEEN CASES. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288307 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1