WECKVISTA OPTICAL PORT 5MMX100MM RIDGES
Report
- Report Number
- 3003898360-2014-00310
- Event Type
- Malfunction
- Date Received
- May 13, 2014
- Date of Event
- April 17, 2014
- Report Date
- April 18, 2014
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE SAMPLE NOT RECEIVED BY MFR.
QN#(B)(4). THE CUSTOMER RETURNED ONE UNIT 405912R WECKVISTA OPTICAL PORT 5MMX100MM RIDGED FOR INVESTIGATION. THE SAMPLE WAS RECEIVED WITHOUT ITS ORIGINAL PACKAGING. THE RETURNED DEVICE WAS VISUALLY EXAMINED WITH AND WITHOUT MAGNIFICATION. VISUAL EXAMINATION OF THE RETURNED UNIT REVEALED THAT THE DEVICE APPEARS TYPICAL. NO DEFECTS OR ANOMALIES WERE OBSERVED AND NO PIECES APPEAR TO BE MISSING (REFERENCE FILES (B)(4)). THE IFU FOR THIS PRODUCT, L05489, WAS REVIEWED AS A PART OF THIS COMPLAINT INVESTIGATION. THE IFU STATES, "BEFORE USING THE WECK VISTA OPTICAL BLADELESS LAPAROSCOPIC ACCESS PORT, CONFIRM THE COMPATIBILITY OF ALL INSTRUMENTS AND ACCESSORIES." THE IFU ALSO INDICATES, "INSERTION AND REMOVAL OF INSTRUMENTS AND LAPAROSCOPE THROUGH THE CANNULA SHOULD BE DONE CAREFULLY TO PREVENT UNINTENTIONAL DAMAGE TO THE INTERNAL SEALS, WHICH MAY COMPROMISE INSUFFLATION. TO ENSURE SMOOTH INSERTION, AVOID INSERTING INSTRUMENTS AT AN ANGLE. ALWAYS ATTEMPT TO INSERT INSTRUMENTATION STRAIGHT DOWN THE CANNULA." A CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME AS THERE WAS NO VISUAL DEFECT OR ANOMALIES FOUND WITH THE RETURNED SAMPLE. OTHER REMARKS: THE REPORTED COMPLAINT OF THE END OF THE PORT BREAKING DURING USE COULD NOT BE CONFIRMED BASED UPON THE SAMPLE RECEIVED. VISUAL INSPECTION OF THE RETURNED SAMPLE DID NOT REVEAL ANY ANOMALIES OR DEFECTS. A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED AS NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER. THE IFU FOR THIS PRODUCT, L05489, WAS REVIEWED AS A PART OF THIS COMPLAINT INVESTIGATION. THE IFU STATES , "BEFORE USING THE WECK VISTA OPTICAL BLADELESS LAPAROSCOPIC ACCESS PORT, CONFIRM THE COMPATIBILITY OF ALL INSTRUMENTS AND ACCESSORIES." THE IFU ALSO INDICATES, "INSERTION AND REMOVAL OF INSTRUMENTS AND LAPAROSCOPE THROUGH THE CANNULA SHOULD BE DONE CAREFULLY TO PREVENT UNINTENTIONAL DAMAGE TO THE INTERNAL SEALS, WHICH MAY COMPROMISE INSUFFLATION. TO ENSURE SMOOTH INSERTION, AVOID INSERTING INSTRUMENTS AT AN ANGLE. ALWAYS ATTEMPT TO INSERT INSTRUMENTATION STRAIGHT DOWN THE CANNULA." THEREFORE, BASED ON THIS, NO VISUAL ISSUES WERE OBSERVED WITH THE RETURNED SAMPLE.
(B)(4). THE CUSTOMER RETURNED ONE UNIT 405912R WECKVISTA OPTICAL PORT 5MMX100MM RIDGED FOR INVESTIGATION. THE SAMPLE WAS RECEIVED WITHOUT ITS ORIGINAL PACKAGING. THE RETURNED DEVICE WAS VISUALLY EXAMINED WITH AND WITHOUT MAGNIFICATION. VISUAL EXAMINATION OF THE RETURNED PORT REVEALED THAT THE TIP OF THE CANNULA IS DAMAGED. A PIECE IS MISSING FROM THE TIP AND THE PLASTIC IS CRACKED AND FRAYED. WHITE DISCOLORATION CAN BE SEEN IN THE PLASTIC MATERIAL NEAR WHERE THE PIECE IS MISSING INDICATING THAT THE CANNULA WAS DAMAGED BY TRAUMA AGAINST A HARD SURFACE. NO OTHER DEFECTS OR ANOMALIES WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED AS NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER. OTHER REMARKS: THE REPORTED COMPLAINT OF A DAMAGED PORT DURING INSERTION OF AN INSTRUMENT WAS CONFIRMED BASED UPON THE SAMPLE RECEIVED. THE RETURNED PORT WAS CONFIRMED TO HAVE A DAMAGED CANNULA. HOWEVER, IT IS UNKNOWN WHAT SIZE OR TYPE OF INSTRUMENTS WAS PASSED THROUGH THE PORT. A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED AS NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER. THE IFU FOR THIS PRODUCT, L05489, WAS REVIEWED AS A PART OF THIS COMPLAINT INVESTIGATION. THE IFU STATES, "BEFORE USING THE WECK VISTA OPTICAL BLADELESS LAPAROSCOPIC ACCESS PORT, CONFIRM THE COMPATIBILITY OF ALL INSTRUMENTS AND ACCESSORIES." THE IFU ALSO INDICATES, "INSERTION AND REMOVAL OF INSTRUMENTS AND LAPAROSCOPE THROUGH THE CANNULA SHOULD BE DONE CAREFULLY TO PREVENT UNINTENTIONAL DAMAGE TO THE INTERNAL SEALS, WHICH MAY COMPROMISE INSUFFLATION. TO ENSURE SMOOTH INSERTION, AVOID INSERTING INSTRUMENTS AT AN ANGLE. ALWAYS ATTEMPT TO INSERT INSTRUMENTATION STRAIGHT DOWN THE CANNULA." THEREFORE, BASED UPON THE TIME OF DISCOVERY AND THE INFORMATION PROVIDED, OPERATIONAL CONTEXT CAUSED OR CONTRIBUTED TO THIS EVENT. A CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME AS THE TIME OF DISCOVERY AND THE DAMAGE OBSERVED INDICATE THAT OPERATIONAL CONTEXT CAUSED OR CONTRIBUTED TO THIS EVENT.
COMPLAINT ALLEGES: THE END OF THE PORT BROKE INTO THE PT ABDOMEN RESULTING IN PIECES FALLING INTO HIS ABDOMEN. MOST PIECES RETRIEVED. THE PT CONDITION IS REPORTED AS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 285628 | WECKVISTA OPTICAL PORT 5MMX100MM RIDGES | OPTICAL PORT | GCJ | TELEFLEX MEDICAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |