FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3912354 · Received July 3, 2014

Report

Report Number
1416980-2014-21440
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, IMPROPERLY DISCONNECTING/RECONNECTING FROM THE SET IS A KNOWN CAUSE OF THIS ALARM. PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE HAS INSTRUCTIONS ON THE STEPS TO PROPERLY DISCONNECT FROM CYCLER DURING AN EMERGENCY AND EXPLAINS HOW TO RECONNECT FOR THERAPY AFTER PROPERLY DISCONNECTING. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON A HOMECHOICE (HC) DEVICE DURING USE. THE PATIENT WAS CONNECTED TO THE HOMECHOICE AT THE TIME OF THE ALARM. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT DISCONNECTED FROM SET IMPROPERLY DURING THERAPY AND RECONNECTED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390660 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE