ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19036
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Report Date
- June 29, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 09/10/2014. DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. REVIEW OF BLACK BOX DATA FOUND LOW BATTERY WARNING ON (B)(6) 2014 AND A PARTIALLY DISCHARGED BATTERY WAS INSTALLED TWO DAYS LATER. THE BATTERY CAP WAS NOT RETURNED AND A TEST BATTERY CAP WAS USED IN THE EVALUATION. THE PUMP POWERED UP AND FUNCTIONED PROPERLY. A REWIND/LOAD/PRIME SEQUENCE AND A 24-HOUR PUMP EXERCISE WERE EXECUTED WITHOUT INCIDENCES. REMOVAL OF THE PUMP CASING DID NOT FIND ANY MOISTURE OR DAMAGES TO THE POWER CIRCUIT. THE REPORTED POWER ISSUE WAS NOT DUPLICATED IN THE INVESTIGATION AND WAS ATTRIBUTED TO A USE ERROR IN THAT A BATTERY WITH LOW VOLTAGE WAS INSTALLED DURING A BATTERY CHANGE. UNRELATED TO THE REPORTED POWER ISSUE, THE BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED BELOW THE GRIP PAD.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. REPORTEDLY, THE PUMP HAD INTERMITTENT POWER ISSUE. IT WAS REPORTED THAT THERE WAS NO DAMAGE TO THE BATTERY COMPARTMENT OR CAP AND THE CAP WAS ABLE TO SECURE PROPERLY ON TO THE PUMP. THE REPORTER NOTED THAT THERE WAS NO MOISTURE/CORROSION IN THE PUMP AND THE ISSUE WAS NOT RESOLVED WITH A NEW BATTERY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391146 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |