FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYTEM 3.8MM

MDR report key: 3912140 · Received May 13, 2014

Report

Report Number
2242352-2014-00548
Event Type
Malfunction
Date Received
May 13, 2014
Report Date
April 16, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, ANCHOR TAB AND THE SEAL WERE NOT RETURNED. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAILURE TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION ((B)(4)) ISSUED FOR THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III FAILED TO LOAD PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285534 HS III PROXIMAL SEAL SYTEM 3.8MM ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25069722