HS III PROXIMAL SEAL SYSTEM 3.8MM
Report
- Report Number
- 2242352-2014-00545
- Event Type
- Malfunction
- Date Received
- May 13, 2014
- Date of Event
- September 26, 2013
- Report Date
- April 16, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC.
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE. THERE WAS SLIGHT EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED INSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, ANCHOR TAB AND SEAL WERE INSIDE THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THIS FAILURE IS POTENTIALLY ATTRIBUTED TO USER TECHNIQUE. IT APPEARS THAT THE SEAL HAD BEEN PREMATURELY DEPLOYED IN THE LOADING DEVICE. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III FAILED TO LOAD PROPERLY. THE SEAL REMAINED IN THE DEVICE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.