FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 3912109 · Received May 13, 2014

Report

Report Number
2242352-2014-00545
Event Type
Malfunction
Date Received
May 13, 2014
Date of Event
September 26, 2013
Report Date
April 16, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE. THERE WAS SLIGHT EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED INSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, ANCHOR TAB AND SEAL WERE INSIDE THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THIS FAILURE IS POTENTIALLY ATTRIBUTED TO USER TECHNIQUE. IT APPEARS THAT THE SEAL HAD BEEN PREMATURELY DEPLOYED IN THE LOADING DEVICE. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III FAILED TO LOAD PROPERLY. THE SEAL REMAINED IN THE DEVICE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285527 HS III PROXIMAL SEAL SYSTEM 3.8MM ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC. HSK-3038 25083696

Patients

Seq Age Sex Outcome Treatment
1