ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19021
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Report Date
- June 28, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE PUMP POWERED UP TO A BLANK DISPLAY WITH AUDITORY AND VIBRATORY FEATURES. FUNCTIONAL TESTING FOR THE BUTTONS WAS NOT PERFORMED DUE TO THE BLANK DISPLAY. THE BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED AT THE THREADS. A LEAK TEST SHOWED A LEAK WHERE THE CRACK WAS LOCATED. PUMP CASING WAS OPENED AND EVIDENCE OF MOISTURE INTRUSION WAS FOUND ON THE DISPLAY AS WELL AS THE DISPLAY CONNECTOR WIRE. REPLACING THE DISPLAY WITH A TEST DISPLAY CORRECTED THE DISPLAY ISSUE. THE REPORTED DISPLAY AND MOISTURE INGRESS ISSUE WAS DUPLICATED IN THE INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY WAS BLANK AND THERE WAS MOISTURE IN THE BATTERY COMPARTMENT. IT WAS REPORTED THAT THE ISSUE WAS NOT RESOLVED WITH PUMP REBOOT OR CONTRAST RESET. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390134 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |