FDA Adverse Event Summary report: N

BRILLIANCE 64

MDR report key: 3911513 · Received May 27, 2014

Report

Report Number
1525965-2014-00111
Date Received
May 27, 2014
Report Date
April 28, 2014
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K033326
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

A PHILIPS FIELD SERVICE ENGINEER (FSE) REPORTED THAT WHILE PERFORMING MAINTENANCE ACTIVITIES, THE "IN" BUTTON ON THE GANTRY CONTROL PANEL WAS STICKING ENGAGED. THE FSE CONFIRMED THERE WAS NO HARM TO A PATIENT, OPERATOR, OR BYSTANDER DUE TO THIS ISSUE. MAINTENANCE ACTIVITIES OCCUR WITH NO PATIENT INVOLVEMENT. THE FSE REPLACED THE RIGHT FRONT AND RIGHT REAR GANTRY CONTROL PANELS TO RESOLVE THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310472 BRILLIANCE 64 JAK PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. 728231

Patients

Seq Age Sex Outcome Treatment
1