TANGO OPTIMO
Report
- Report Number
- 9610824-2014-00049
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- May 31, 2014
- Report Date
- July 9, 2014
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMBH
- Product Code
- KSZ
- PMA / PMN Number
- BK080013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS IS OUR INITIAL REPORT ON THIS INCIDENT.
THE STATED DATE OF EVENT IN OUR FOLLOW-UP REPORT #01 WAS INCORRECT. IT SHOULD HAVE BEEN (B)(6) 2014 INSTEAD OF (B)(6) 2014.
THIS IS OUR FINAL REPORT ON THIS INCIDENT.
THE CUSTOMER REPORTED THAT HE MISSED AN ANTI-E AND ANTI-S WHEN TESTING WITH BIOTESTCELL 3 ON TANGO OPTIMO. THE CUSTOMER REPORTED THAT THE IMAGES APPEARED VISUALLY POSITIVE AND THEREFORE THE NEGATIVE RESULTS OF TANGO OPTIMO WERE EDITED BY TECHNICIAN. THE CUSTOMER HAS NEITHER RETURNED THE SAMPLES THAT HAD CAUSED A FALSE NEGATIVE TEST RESULT NOR THE SUPPOSEDLY DEFECTIVE PRODUCT NOR THE IMAGES OF TANGO OPTIMO. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THE RETAINED TANGO OPTIMO SAMPLE WITH NEGATIVE AND POSITIVE CONTROL AND SOME ANTIBODIES, E.G. ANTI-E. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE NEGATIVE REACTION. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THAT THE ALLEGEDLY DEFECTIVE LOT OF BIOTESTCELL 3 FUNCTIONS CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. THE AFFECTED TANGO OPTIMO WAS INSPECTED. ITS CAMERA VALUES WERE FOUND TO BE OUT OF RANGE AND HAVE BEEN ADJUSTED TO THE VALID RANGE. AFTER THIS SERVICE INTERVENTION, THE CUSTOMER FEELS COMFORTABLE WITH THE INSTRUMENT PERFORMANCE NO FURTHER ISSUES WERE REPORTED.
THE CUSTOMER REPORTED THAT HE MISSED AN ANTI-E AND ANTI-S WHEN TESTING WITH BIOTEST CELL 3 ON TANGO OPTIMO. THE CUSTOMER REPORTED THAT THE IMAGES APPEARED VISUALLY POSITIVE AND THEREFORE THE NEGATIVE RESULTS OF TANGO OPTIMO WERE EDITED BY TECHNICIAN. THE CUSTOMER HAS NEITHER RETURNED THE SAMPLES THAT HAD CAUSED A FALSE NEGATIVE TEST RESULT NOR THE SUPPOSEDLY DEFECTIVE PRODUCT NOR THE IMAGES OF TANGO OPTIMO. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THE RETAINED TANGO OPTIMO SAMPLE WITH NEGATIVE AND POSITIVE CONTROL AND SOME ANTIBODIES, E.G. ANTI-E. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE NEGATIVE REACTION. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THAT THE ALLEGEDLY DEFECTIVE LOT OF BIOTESTCELL 3 FUNCTIONS CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. THE AFFECTED TANGO OPTIMO WAS INSPECTED. ITS CAMERA VALUES WERE FOUND TO BE OUT OF RANGE AND HAVE BEEN ADJUSTED TO THE VALID RANGE. AFTER THIS SERVICE INTERVENTION, THE CUSTOMER FEELS COMFORTABLE WITH THE INSTRUMENT PERFORMANCE NO FURTHER ISSUES WERE REPORTED.
THE CUSTOMER REPORTED THAT HE MISSED AN ANTI-E AND ANTI-S WHEN TESTING WITH BIOTESTCELL 3 ON TANGO OPTIMO. THE CUSTOMER REPORTED THAT THE IMAGES APPEARED VISUALLY POSITIVE AND THEREFORE THE NEGATIVE RESULTS OF TANGO OPTIMO WERE EDITED BY TECHNICIAN. THE CUSTOMER HAS NEITHER RETURNED THE SAMPLES THAT HAD CAUSED A FALSE NEGATIVE TEST RESULT NOR THE SUPPOSEDLY DEFECTIVE PRODUCT NOR THE IMAGES OF TANGO OPTIMO. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THE RETAINED TANGO OPTIMO SAMPLE WITH NEGATIVE AND POSITIVE CONTROL AND SOME ANTIBODIES, E.G. ANTI-E. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE NEGATIVE REACTION. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THAT THE ALLEGEDLY DEFECTIVE LOT OF BIOTESTCELL 3 FUNCTIONS CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. THE AFFECTED TANGO OPTIMO WAS INSPECTED. ITS CAMERA VALUES WERE FOUND TO BE OUT OF RANGE AND HAVE BEEN ADJUSTED TO THE VALID RANGE. WE ARE WAITING FOR A FOLLOW-UP REPORT ON THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 389174 | TANGO OPTIMO | AUTOMATED BLOOD BANK ANALYSER SYSTEM | KSZ | BIO-RAD MEDICAL DIAGNOSTICS GMBH | 8416021-00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | AHG, LOT 8343150-03| MLB2, LOT 8344050-00| MLB2, LOT 8344050-00| AHG, LOT 8343150-03| BIOTESTCELL 3, LOT 8416021-00, EXP. 06/09/2014| MLB2, LOT 8344050-00| TANGO OPTIMO, # (B)(4)| AHG, LOT 8343150-03| BIOTESTCELL 3, LOT 8416021-00, EXP. 06/09/2014 |