FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3911148 · Received July 2, 2014

Report

Report Number
9611451-2014-00586
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 9, 2014
Report Date
June 16, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED.RESULTS: VISUAL INSPECTION REVEALED A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. THE SURFACE AT THE BREAK WAS ROUGH. THE DAMAGE APPEARED TO BE DUE TO THE TUBE BEING PULLED AWAY FROM THE CHAMBER, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR PUT UNDER TENSION.A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 131012.CONCLUSION: WE WERE UNABLE TO DEFINITIVELY DETERMINE THE CAUSE OF THE FEEDSET BREAK, HOWEVER OUR PREVIOUS INVESTIGATIONS INTO SIMILAR COMPLAINTS HAVE SHOWN THAT THIS DAMAGE CAN BE CAUSED BY THE OVERSTRETCHING OF THE TUBING WHEN ATTACHING IT TO A WATER BAG.WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THE CUSTOMER HAS CONFIRMED THAT THE SUBJECT CHAMBER WAS IN USE FOR ONE WEEK BEFORE A LEAK WAS OBSERVED, WHICH INDICATES THAT THE CHAMBER BECAME DAMAGED DURING USE.THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 AUTOFEED HUMIDIFICATION CHAMBERS STATE THE FOLLOWING:"SET APPROPRIATE VENTILATOR ALARMS.""PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."(B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE WATER FEEDSET OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER STARTED TO LEAK AT THE CHAMBER DOME CONNECTION AFTER ONE WEEK OF USE.NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385987 VENTED AUTOFEED CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 131012

Patients

Seq Age Sex Outcome Treatment
1