FDA Adverse Event
Injury
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 3911131
·
Received July 2, 2014
Report
- Report Number
- 2938836-2014-12971
- Event Type
- Injury
- Date Received
- July 2, 2014
- Date of Event
- May 30, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC FOR ROUTINE FOLLOW UP. SEVERAL EPISODE DIAGNOSED AS VF OCCURRED. NOISE WAS OBSERVED ON THE RV LEAD. THE LEAD WAS CAPPED AND REPLACED. THE PATIENT WAS IN GOOD CONDITION FOLLOWING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386202 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 1580/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Required Intervention |