DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 3911072
·
Received July 2, 2014
Report
- Report Number
- 2938836-2014-12998
- Event Type
- Injury
- Date Received
- July 2, 2014
- Date of Event
- April 30, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.
Description of Event or Problem · 1
THE PATIENT CAME INTO THE HOSPITAL AFTER REPORTEDLY RECEIVING A VIBRATORY ALERT. DEVICE INTERROGATION REVEALED DECREASED SENSING. X-RAY REVEALED LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385828 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7122Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention |