FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 3911072 · Received July 2, 2014

Report

Report Number
2938836-2014-12998
Event Type
Injury
Date Received
July 2, 2014
Date of Event
April 30, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.

Description of Event or Problem · 1

THE PATIENT CAME INTO THE HOSPITAL AFTER REPORTEDLY RECEIVING A VIBRATORY ALERT. DEVICE INTERROGATION REVEALED DECREASED SENSING. X-RAY REVEALED LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385828 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention