RIATA ST ACTIVE FIXATION
MDR report key: 3911062
·
Received July 2, 2014
Report
- Report Number
- 2938836-2014-12992
- Event Type
- Injury
- Date Received
- July 2, 2014
- Date of Event
- May 2, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GB
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. A PARTIAL LEAD WITH THE DISTAL PORTION MEASURING 41.0CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 13.5-14.0CM FROM THE DISTAL END. THE ETFE COATING WAS INTACT AT THIS LOCATION. EXTERNAL INSULATION ABRASION WAS NOTED AT 7.0-8.0CM FROM THE DISTAL END. THE ETFE COATING WAS ABRADED AT THIS LOCATION. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO NOISE. LEAD WAS EXPLANTED. PATIENT WAS WELL AFTER PROCEDURE.