RIATA ST ACTIVE FIXATION

MDR report key: 3911062 · Received July 2, 2014

Report

Report Number
2938836-2014-12992
Event Type
Injury
Date Received
July 2, 2014
Date of Event
May 2, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. A PARTIAL LEAD WITH THE DISTAL PORTION MEASURING 41.0CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 13.5-14.0CM FROM THE DISTAL END. THE ETFE COATING WAS INTACT AT THIS LOCATION. EXTERNAL INSULATION ABRASION WAS NOTED AT 7.0-8.0CM FROM THE DISTAL END. THE ETFE COATING WAS ABRADED AT THIS LOCATION. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO NOISE. LEAD WAS EXPLANTED. PATIENT WAS WELL AFTER PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385825 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention