FDA Adverse Event Injury Summary report: N

UNIFY CRT-D, DF-4 CONNECTOR

MDR report key: 3911045 · Received July 2, 2014

Report

Report Number
2938836-2014-13031
Event Type
Injury
Date Received
July 2, 2014
Date of Event
May 20, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM. POST-SENSED T-WAVE OVERSENSING WAS OBSERVED ON STORED EGM, AND THE DEVICE DELIVERED INAPPROPRIATE THERAPY. THE DEVICE WAS REPROGRAMMED AND REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385923 UNIFY CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3231-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention