FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3911043 · Received July 2, 2014

Report

Report Number
2938836-2014-13058
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
May 1, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 14.5CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DEVICE CHANGE OUT DUE TO EOL, EXTERNALIZED CONDUCTORS WERE NOTED ON THE RV LEAD. NO ELECTRICAL ANOMALIES WERE DETECTED. THE LEAD WAS CAPPED AND REPLACED. THE PATIENT WAS IN GOOD CONDITION FOLLOWING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
386171 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1 79 YR