VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2014-00584
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- May 15, 2014
- Report Date
- June 3, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RECEIVED AT FISHER & PAYKEL HEALTHCARE PS159133 FOR EVALUATION. RESULTS: VISUAL INSPECTION REVEALED THAT THE CHAMBER DOME WAS CRACKED UNDERNEATH THE CHAMBER PORT. THE PRINT ABOVE THE CRACK AND BENEATH THE HINGE BRACKET WAS SLIGHTLY SMEARED. A LOT CHECK REVEALED NO OTHER SIMILAR COMPLAINTS FOR LOT DATE 130205. CONCLUSION: THE RESIDUE AROUND THE CRACK SUGGESTS THAT THE DAMAGE WAS CAUSED BY THE CHAMBER COMING INTO CONTACT WITH A SOLUTION CONTAINING ALCOHOL WHICH RESULTED IN ENVIRONMENTAL STRESS CRACKING OF THE CHAMBER DOME. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 200 CMH2O FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: "DO NOT SOAK, WASH, STERILIZE, OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS, OR HAND SANITIZERS." "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "MAXIMUM OPERATING PRESSURE: 8 KPA".
A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN MR290 AUTOFEED HUMIDIFICATION CHAMBER LEAKED DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386151 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 130205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |